Retinal thickness analysis with time and spectral-domain Optical Coherence Tomography. Cross-platform interchangeability of manual measurements
Keywords:
Time-domain OCT, Spectral-domain OCT, Stratus, Cirrus, Retinal Thickness, Retinal layers, OCT ComparisonAbstract
Abstract. Background and aim of the work: retinal thickness values obtained by automated analysis with new spectral-domain optical coherence tomography (OCT) devices exceed those measured by old time-domain OCTs. Aim of the present work is to assess reproducibility and comparability of manual measurements performed on both time-domain and spectral-domain OCT scans. Methods: 6 eyes were scanned using Stratus OCT3 and Cirrus hd-OCT. Raw exported images were analyzed by ordinary computer software; multiple manual measurements of retinal thickness were taken at different eccentricities. Stratus Retinal Thickness (SRT) was measured from Internal Limiting Membrane (ILM) to the photoreceptors Internal-Outer Segment interface (IS/OS), while two series of measurements were performed in Cirrus images: CRT = Cirrus-RT (from ILM to the Retinal Pigment Epitelium, RPE) and cCRT = corrected-CRT (from ILM to IS/OS). Measurements were repeated twice in two eyes and reproducibility was assessed by Intra-Class correlation (ICC) and Coefficient of Variation (CV). Bland-Altman plots, paired t-test and ICC were used for comparative analysis between Stratus and Cirrus measurements. Results: Mean SRT, CRT and cCRT values ± SD were respectively 244.75 ± 34.78 mm, 275.78 ± 34.36 mm and 244.95 ± 33.78 mm. Paired t-test resulted in p<0.0001 comparing SRT and CRT series, versus p=0.6544 between SRT and cCRT series. ICC was 0.65 between SRT and CRT and 0,94 between SRT-cCRT. Reproducibility of the measurements was excellent (CV=2,13% for Stratus and 1,62% for Cirrus; ICC=0,994 for both devices). Discussion: while a systematic error affects comparison of Stratus and Cirrus macular thickness maps, manual linear measurements result interchangeable, hence allowing comparison of images acquired with either OCT systems.Downloads
Published
Issue
Section
License
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Transfer of Copyright and Permission to Reproduce Parts of Published Papers.
Authors retain the copyright for their published work. No formal permission will be required to reproduce parts (tables or illustrations) of published papers, provided the source is quoted appropriately and reproduction has no commercial intent. Reproductions with commercial intent will require written permission and payment of royalties.