Retinal thickness analysis with time and spectral-domain Optical Coherence Tomography. Cross-platform interchangeability of manual measurements

Retinal thickness analysis with time and spectral-domain Optical Coherence Tomography. Cross-platform interchangeability of manual measurements

Authors

  • A. Neri
  • E. Delfini
  • C. Casubolo, et al.

Keywords:

Time-domain OCT, Spectral-domain OCT, Stratus, Cirrus, Retinal Thickness, Retinal layers, OCT Comparison

Abstract

Abstract. Background and aim of the work: retinal thickness values obtained by automated analysis with new spectral-domain optical coherence tomography (OCT) devices exceed those measured by old time-domain OCTs. Aim of the present work is to assess reproducibility and comparability of manual measurements performed on both time-domain and spectral-domain OCT scans. Methods: 6 eyes were scanned using Stratus OCT3 and Cirrus hd-OCT. Raw exported images were analyzed by ordinary computer software; multiple manual measurements of retinal thickness were taken at different eccentricities. Stratus Retinal Thickness (SRT) was measured from Internal Limiting Membrane (ILM) to the photoreceptors Internal-Outer Segment interface (IS/OS), while two series of measurements were performed in Cirrus images: CRT = Cirrus-RT (from ILM to the Retinal Pigment Epitelium, RPE) and cCRT = corrected-CRT (from ILM to IS/OS). Measurements were repeated twice in two eyes and reproducibility was assessed by Intra-Class correlation (ICC) and Coefficient of Variation (CV). Bland-Altman plots, paired t-test and ICC were used for comparative analysis between Stratus and Cirrus measurements. Results: Mean SRT, CRT and cCRT values ± SD were respectively 244.75 ± 34.78 mm, 275.78 ± 34.36 mm and 244.95 ± 33.78 mm. Paired t-test resulted in p<0.0001 comparing SRT and CRT series, versus p=0.6544 between SRT and cCRT series. ICC was 0.65 between SRT and CRT and 0,94 between SRT-cCRT. Reproducibility of the measurements was excellent (CV=2,13% for Stratus and 1,62% for Cirrus; ICC=0,994 for both devices). Discussion: while a systematic error affects comparison of Stratus and Cirrus macular thickness maps, manual linear measurements result interchangeable, hence allowing comparison of images acquired with either OCT systems.

Downloads

Published

01-02-2012

Issue

Section

ORIGINAL ARTICLES

How to Cite

1.
Neri A, Delfini E, Casubolo, et al. C. Retinal thickness analysis with time and spectral-domain Optical Coherence Tomography. Cross-platform interchangeability of manual measurements. Acta Biomed [Internet]. 2012 Feb. 1 [cited 2024 Jul. 27];82(3):244-50. Available from: https://mattioli1885journals.com/index.php/actabiomedica/article/view/1735